Understanding how to compare results taken with the OneTouch Ultra Plus Flex™ meter to laboratory results including guidelines for obtaining an accurate meter to lab comparison and information for testing at the lab.
Product Guide
USA
OneTouch Ultra Plus Flex™ meter
English
0.0
Article Details
Comparing Meter Results To Laboratory Results:
Results obtained from the OneTouch Ultra Plus Flex™ Meter and laboratory tests are reported in plasma-equivalent units. However, your meter result may differ from your lab result due to normal variation. A result from your OneTouch Ultra Plus Flex™ Meter is considered accurate when it is within 15 mg/dL of a laboratory method when the glucose concentration is lower than 75 mg/dL and within 15% of a laboratory method when the glucose concentration is 75 mg/dL or higher.
Meter results can be affected by factors that do not affect lab results in the same way. Specific factors that may cause your meter result to vary from your lab result may include:
You have eaten recently. This can cause a result from fingertip testing to be up to 70 mg/dL higher than a lab test using blood drawn from a vein.1
1Sacks, D.B.: "Carbohydrates." Burtis, C.A., and Ashwood E.R. (ed.), Tietz Textbook of Clinical Chemistry, Philadelphia: W.B. Saunders Company (1994), 959.
Your hematocrit is above 60% or below 20%.
You are severely dehydrated.
For additional information, refer to the OneTouch Ultra Plus™ Test Strip Insert.