Testing With Control Solution:
Control Solution Testing Precautions:
- OneTouch Ultra Plus™ Control Solution is used to check that the meter and test strips are working together properly and that the test is performing correctly. (Control solution is available separately.)
- Note:
- Use only OneTouch Ultra Plus™ Level 3 Control Solution or OneTouch Ultra Plus™ Level 4 Control Solution with your OneTouch Ultra Plus Flex™ Meter. Either level can be used to check your system.
- When you first open a new vial of control solution, record the discard date on the vial label. Refer to the control solution insert or vial label for instructions on determining the discard date.
- Tightly close the cap on the control solution vial immediately after use to avoid contamination or damage.
- For more information on testing:
- CAUTION:
- Do Not swallow or ingest control solution.
- Do Not apply control solution to the skin, eyes, ears or nose as it may cause irritation.
- Do Not use control solution after the expiration date (printed on the vial label) or the discard date, whichever comes first, or your results may be inaccurate.
When To Do A Control Solution Test:
- When you open a new vial of test strips.
- If you suspect that the meter or test strips are not working properly.
- If you have had repeated unexpected blood glucose readings.
- If you drop or damage the meter.
Preparing Your Meter For A Control Solution Test:
1. Insert a test strip to turn the meter on
2. Wait for the flashing blood drop icon
to appear on the display
3. Press and hold ↑ or ↓ until the control solution icon
appears on the display
Preparing The Control Solution:
1. Before removing the cap, shake the vial gently
2. Remove the vial cap and place it on a flat surface with the top of the cap pointing up
3. Squeeze the vial to discard the first drop
4. Wipe both the tip of the control solution vial and the top of the cap with a clean, damp tissue or cloth
5. Squeeze a drop into the small well on the top of the cap or onto another clean, non-absorbent surface
Applying The Control Solution:
1. Hold the meter so that the side edge of the test strip is at a slight angle to the drop of control solution
2. Touch the channel on the side of the test strip to the control solution
3. Wait for the channel to fill completely