Understanding test procedure specifics when testing with the OneTouch® Ultra® Control Solution.
Product Guide
Global
OneTouch® Delica® Plus lancets; OneTouch® Delica® Plus lancing device; OneTouch Ultra® control solution; OneTouch Ultra® test strips
English
3.7
Article Details

Test Procedure:
 

Intended Use:

  • For use with the OneTouch® Ultra® Family of Meters to check that the meter and test strips are working together properly and that you are performing the test correctly.

Do a Control Solution Test:

  • Carefully read the instructions in your Owner’s Booklet or User Guide as well as articles provided on when to perform a control solution test.
  • The OneTouch® Ultra® Control Solution test result should be within the expected range printed on the test strip vial if the meter and test strips are working properly and you performed the test correctly.

Test Procedure:

  • Materials provided:
    • OneTouch® Ultra® Control Solution (vial with blue cap).
  • Materials required but not provided:
    • Your meter
    • Owner’s Booklet, User Guide or articles provided
    • OneTouch® Ultra® Test Strips
  • Carefully read the instructions in your Owner’s Booklet or User Guide as well as articles provided on how to perform a control solution test.
  • Some OneTouch® meters require coding. See the instructions that came with your system for directions on matching the code on the meter with the code on the test strip vial. If your meter requires coding, accurate control solution test results depend upon the meter being properly coded to match the number on the test strip vial.

Important Safety Information:

  • Use OneTouch® Ultra® Control Solution with the OneTouch® Ultra® Family of Meters.
  • Write the discard date (date opened plus 3 months) in the space provided on the vial label when you first open it.
  • Do Not use after the expiration (printed on vial label) or discard date, whichever comes first.
  • Control solution tests must be done at room temperature (68–77°F).
  • CAUTION:
    • Do Not swallow control solution; it is not for human consumption.
    • Do Not apply control solution to the skin or eyes as it may cause irritation.