OneTouch Verio® – Test With Control Solution

« Go Back
Information
OneTouch Verio® – Test With Control Solution
OT-Verio-Test-with-Control-Solution
Instructions for performing a control solution test using the OneTouch Verio® meter including notes, cautions, when to do a control solution test, understanding why out-of-range results may appear and how to clean the control solution cap.
Product Guide
Global
OneTouch Verio® control solution; OneTouch Verio® meter
English
4.6
Article Details
Test With Control Solution:
 
  • OneTouch Verio® Control Solution is used to check that the meter and test strips are working together properly and that the test is performing correctly. (Control solution is available separately.)
  • Note:
    • Use only OneTouch Verio® Level 3 Control Solution or OneTouch Verio® Level 4 Control Solution with your OneTouch Verio® Meter. Either level can be used to check your system.
    • When you first open a new vial of control solution, record the discard date on the vial label. Refer to the control solution insert or vial label for instructions on determining the discard date.
    • Tightly close the cap on the control solution vial immediately after use to avoid contamination or damage.
  • CAUTION:
    • Do Not swallow or ingest control solution.
    • Do Not apply control solution to the skin or eyes as it may cause irritation.
    • Do Not use control solution after the expiration date (printed on the vial label) or the discard date, whichever comes first, or your results may be inaccurate.
Do A Control Solution Test:
  • Whenever you open a new vial of test strips.
  • If you suspect that the meter or test strips are not working properly.
  • If you have had repeated unexpected blood glucose results.
  • If you drop or damage the meter.

Performing A Control Solution Test:
 
1. Insert a test strip to turn the meter on
 
  • Wait for the Apply Blood screen to appear on the display.
  • Note: The same Apply Blood screen that appears during a blood glucose test also appears during a control solution test.
2. Prepare the control solution
  • Remove the vial cap and place it on a flat surface with the top of the cap pointing up.
  • Squeeze the vial to discard the first drop.
 image.png
  • Wipe both the tip of the control solution vial and the top of the cap with a clean, damp tissue or cloth.
  • Then, squeeze a drop into the small well on the top of the cap or onto another clean, non-absorbent surface.
3. Apply the control solution
  • Hold the meter so that the side edge of the test strip is at a slight angle to the drop of control solution.
  • Touch the channel on the side of the test strip to the control solution. Wait for the channel to fill completely.
4. Read your result
  • The meter will count down from 5 to 1. Your result is displayed along with the date, time, unit of measure, and image.png Control solution.
  • The meter automatically marks the result as a control solution test.
image.png
  • CAUTION: If the words Control solution do not appear on the screen, this result will be included in your averages and your averages will change too. Repeat the test with a new test strip. If the problem persists, contact OneTouch® Customer Care. 
5. Check if the result is in range
  • Each vial of test strips has both OneTouch Verio® Level 3 Control Solution and OneTouch Verio® Level 4 Control Solution ranges printed on its label. Compare the result displayed on the meter to either the OneTouch Verio® Level 3 Control Solution or OneTouch Verio® Level 4 Control Solution range printed on the test strip vial, depending on the type of control solution you used.
image.png
  • Out-of-range results may be due to:
    • Not following the instructions above in this article.
    • Control solution is contaminated, expired, or past its discard date.
    • Test strip or test strip vial is damaged, expired, or past its discard date.
    • Meter, test strips and/or control solution were not all at the same temperature when the control solution test was performed.
    • A problem with the meter.
    • Dirt or contamination in the small well on the top of the control solution cap (see Step 2 instructions above in this article).
  • After getting a control solution result:
    • Once you have a control solution result, you may:
      • Press and hold image.png to return to the Main Menu.
      • Orpress and hold OK for several seconds until the meter turns off. The meter will also automatically turn off if left alone for two minutes.
6. Cleaning
  • Clean the top of the control solution cap with a clean, damp tissue or cloth.
  • Control solution results can be seen when reviewing past results, but are not included in result averages.
  • CAUTION:
    • If you continue to get control solution results that fall outside the range printed on the test strip vial, Do Not use the meter, test strips, or control solution. Contact OneTouch® Customer Care. 
    • The control solution ranges printed on the test strip vial are for control solution tests only and are not recommended ranges for your blood glucose level.
AW 06908604A
 